Studio size D×W×H: 950×950×850mm
Appearance size D×W×H: 1350×1400×2100mm
Temperature range: 0℃~80℃ (not within the measurement range)
Temperature deviation: ±2℃ (without lighting)
Humidity range: 30~98%R.H Humidity deviation: ±2.5%R.H (humidity≥75%R.H) ±3%R.H(humidity≤75%R.H) (without light)
Type of light source: Water-cooled full solar spectrum long arc xenon lamp (average service life is about 1200 hours)
Rain time: 1~9999 minutes, adjustable
Rain cycle: 1~240 minutes, adjustable interval (break)
Spectral wavelength: 290nm~1200nm
Xenon lamp power: 1CW, 6KW (lifetime: 1600 hours)
Xenon lamp power: 6KW
Light source wavelength range: 290-1200nm
Heating power: about 6KW
Humidification power: about 4KW
Light source irradiance: Spectrum 290-1200nm 300-1200w/m2 (continuously adjustable); The energy compensation of the xenon lamp is realized by adjusting the voltage and current.
Irradiation area: about 7500cm2
Distance from arc center to sample holder: 350~380mm
Rotation speed of sample rotating basket: ≥1r/min (stepless speed regulation)
Water spray cycle: Water spray time: 0~99h59min, continuously adjustable; stop spray time: 0~99h59min, continuously adjustable.
Illumination time: continuous programmable control
Blackboard temperature: 65℃±3℃ adjustable
Cooling method: mechanical cooling and air cooling
Total power and power supply: about 15KW; 380V±10% 50HZ
The main technical parameters of the above products are measured at room temperature below 26℃, relative humidity ≤85%R.H, and no-load working room, after two hours of constant temperature and humidity.
Temperature and humidity operation control system
Time Controller: Imported programmable time computer integrated controller
Accuracy range: Setting accuracy: temperature ±0.1℃, humidity ±2%R.H, indication accuracy: temperature ±0.1℃, humidity ±1%R.H
Temperature and humidity sensor: platinum resistance. PT100Ω/MV
Blackboard temperature: Bimetallic blackboard thermometer
Water supply system: Humidification water supply adopts automatic control, xenon lamp cooling water circulation
Temperature Controller: Imported from South Korea, ternary microcomputer touch screen temperature and humidity integrated controller
Circulation system: temperature-resistant and low-noise air-conditioning motor. Multi-blade centrifugal wind wheel
Control system: Mainly composed of operation panel, central controller, various key switches, status indicator lights, executive appliances, etc.
Outer box:
The outer box is formed by CNC machine tools, and the outer shell is formed by high-quality cold-rolled steel plate, and the surface is electrostatically sprayed after pickling and phosphate treatment;
Door and door buckle:
The door seal adopts double-layer refined silicon rubber, so that there is no aging and hardening phenomenon under high and low temperature conditions, and the sealing is more reliable;
The test box is a single door (embedded door handle), with a non-reactional handle;
The door is equipped with an observation window, anti-frost device and switchable lighting lamp;
Observation window:
Size: 25 cm W × 30 cm H;
Radiation-proof hollow glass observation window;
The electric control part of the test box:
This part is the core part of the test chamber;
The main switch of the power supply is an air pressure switch, which controls the overall circuit part;
All line parts have strict line markings for easy maintenance;
Safety protection measures for the electronic control part: When the control system detects the following protection devices (single row), it should be able to automatically cut off the whole system and give an alarm;
Studio over-temperature protection;
Power supply phase loss and leakage protection;
Overcurrent protection of blower motor;
Refrigerator overpressure/overload protection;
Short circuit/overload protection for heaters and humidifiers; Fuse protection;
Working part of the test chamber: Compressor condensation part:
Refrigeration system: The core of the refrigeration system is the compressor. In this solution, we use 2 French Taikang fully enclosed compressors to ensure the cooling requirements of the studio.
The design of the refrigeration system applies energy adjustment technology. An effective treatment method can ensure that the energy consumption and cooling capacity of the refrigeration system can be adjusted under the normal operation of the refrigeration unit, so that the operation cost and failure of the refrigeration system The rate dropped to a more economical state.
Refrigeration evaporator: The evaporator is located in the interlayer of the air duct at one end of the test box, and it is forced to ventilate by a blower motor to quickly exchange heat.
Refrigeration system cooling method: forced air cooling;
Refrigerant: R404a; (environmental friendly refrigerant)
Heating part: Heating: Direct heating in the test box;
The military-grade stainless steel finned heating tube is adopted, and the heating rate is stable, which can keep the temperature in the box without sharp changes. It has the advantages of corrosion resistance, high and low temperature resistance, no leakage, and easy installation; Using SSR (solid state relay) output control, it has the advantages of no noise, reliable work, long life, strong anti-interference ability, fast switching speed, small external interference, and convenient use;
Rain shower system:
The rain shower system is mainly composed of nozzles, solenoid valves, water supply sources and connecting hoses.